pH Meters

Bante902P Portable pH/Conductivity/TDS/Salinity Meter

Multiparameter portable pH/conductivity meter contains 6 measurement modes, the meter is suitable for measuring the pH, ORP, conductivity, total dissolved solids (TDS), salinity and resistivity.

pH Mode:
• Multiparameter portable pH/conductivity meter is equipped with a backlit LCD display.
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Automatic electrode diagnosis shows the pH slope and offset.

Portable pH Conductivity TDS Meter

ORP Mode:
• 1 point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt modes ensure the reliable ORP measurements.

Conductivity/TDS/Salinity/Resistivity Modes:
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear
  and pure water compensations, seawater and practical salinity measurement modes.

Portable pH Conductivity TDS Meter

Other Features:
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-Read function senses and locks the measurement endpoint.
• Calibration due alarm prompts user to calibrate the meter regularly.
• Auto-Power Off effectively conserves battery life.
• Setup menu allows setting the number of calibration points, stability criteria, temperature
  unit, date and time, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface is easy to transfer data to PC.
• Multi-mode power scheme (batteries, power adapter and USB port) ensures that using the
  meter smoothly.

   Model


Bante902P

   pH



   Range


-2.000 to 20.000pH

   Resolution


0.001, 0.01, 0.1pH, selectable

   Accuracy


±0.002pH

   Calibration Points


1 to 5 points

   pH Buffer Options


USA, NIST, DIN, 2 custom buffers

   ORP



   Range


±1999.9mV

   Resolution


0.1, 1mV, selectable

   Accuracy


±0.2mV

   Calibration Points


1 point

   Measurement Modes


Relative or absolute mV

   Conductivity



   Range


0.01 to 20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm

   Resolution


0.001, 0.01, 0.1, 1

   Accuracy


±0.5% F.S.

   Calibration Points


1 to 5 points

   Calibration Solutions


10µS, 84µS, 1413µS, 12.88mS, 111.8mS

   Temperature Coefficient


Linear (0.0 to 10.0%/°C), pure water

   Reference Temperature


20/25°C

   Cell Constant  


K=0.1, 1, 10 or custom

   TDS



   Range


0.00 to 10.00, 100.0, 1000ppm, 10.00, 200.0ppt

   Resolution


0.01, 0.1, 1

   Accuracy

±1% F.S.

   TDS Factor

0.1 to 1.0 (default 0.5)

   Salinity



   Range


0.00 to 80.00ppt, 0.00 to 42.00psu

   Resolution


0.01

   Accuracy

±1% F.S.

   Resistivity



   Range


0.00 to 20.00MΩ

   Resolution


0.01

   Accuracy


±1% F.S.

   Temperature



   Range


0 to 105°C/32 to 221°F

   Resolution


0.1°C/0.1°F

   Accuracy


±0.5°C/±0.9°F

   Offset Calibration


1 point

   Calibration Range


Reading ±10°C

   General Specifications



   Temperature Compensation


0 to 100°C/32 to 212°F, manual or automatic

   Stability Criteria

Low or high

   Calibration Due Alarm


1 to 31 days or off

   Auto-Power Off

10, 20 or 30 minutes after last key pressed

   Memory

Stores up to 500 data sets

   Output

USB communication interface

   Connector

BNC, 6-pin nimi-DIN, 3.5mm jack socket

   Display

Custom LCD (80×60mm)

   Power

3×1.5V AA batteries or DC5V power adapter

   Battery Life

Approximately 150 hours (turn off the backlight)

   Dimensions


170(L)×85(W)×30(H)mm

   Weight


300g

   Order Code

Accessories

   Bante902P-CN

E201-BNC pH electrode

CON-1 conductivity electrode

TP-10K temperature probe

pH4.01/7.00/10.01 buffer sachets

Conductivity standards 84µS/cm, 1413µS/cm, 12.88mS/cm

Carrying case

   Bante902P-UK

   

P11 glass pH electrode

K10 conductivity electrode

TP-10K temperature probe

pH4.01/7.00/10.01 buffer sachets

Conductivity standards 84µS/cm, 1413µS/cm, 12.88mS/cm

Carrying case

Items Description
P11 pH Electrode For the general applications
P11-LiCl pH Electrode For the non-aqueous samples
P11-HA pH Electrode For the high alkalines samples
P11-NA pH Electrode For the biofuels
P12 pH Electrode For the sample in the test tube
P13 pH Electrode For the micro-volume samples
P15 pH Electrode For the low conductivity samples
P16 pH Electrode For the Tris buffers
P18 pH Electrode For the slurries and soils
P19 pH Electrode For the semisolids
P21 pH Electrode For the colloids
P22 pH Electrode For the high temperature samples
501 ORP Electrode For the sample with strong redox potential
502 ORP Electrode For the sample with weak redox potential
CON-0.1 Conductivity Electrode For the pure water
CON-1 Conductivity Electrode For the general water
CON-10 Conductivity Electrode For the high conductivity water
TP-10K Temperature Probe For the temperature measurement
pH Buffer Solutions 480ml/bottle
Conductivity Standard Solutions 480ml/bottle

   Order Code

Carton Dimensions (L×W×H)

Weight    

   Bante902P-CN

340×290×80mm

1.5kg    

   Bante902P-UK

340×290×80mm

1.5kg